ISSN : 1598-6616(Print)
ISSN : 2287-6731(Online)
Journal of Korean Society of Odor Research and Engineering Vol.6 No.4 pp.239-245
Abstract
Analytical methods for odorous butanol compound were developed using canister sampling method and solid adsorbent sampling method. Stability of butanol gas component was evaluated for enhancing accuracy of measurement and to avoid a problem of its poor stability due to its polar character. Quality control method of the analysis was also examined. Gaseous standard gas containing BuOHl and MEK was prepared using canister for evaluation of stability and analytical linearity. BuOH showed a poor stability and poor linearity compared to MEK. However, when the same VOC standard was prepared using adsorbent tube, the stability and linearity of BuOH were much better compared to canister method. Therefore, sampling for the measurement of BuOH in air using adsorbent tube can give better result compared to canister method.
공기중 Butanol 악취물질의 측정법 확립
허귀석,김용두,유연미,김미언,전은주
초록