Journal Search Engine
Download PDF Export Citation Korean Bibliography PMC Previewer
ISSN : 1598-6616(Print)
ISSN : 2287-6731(Online)
Journal of Korean Society of Odor Research and Engineering Vol.9 No.3 pp.133-141
 

GC/NPD 방식에 기초한 아민류 악취물질의 분석특성 비교

Abstract

In this study, calibration characteristics of odorous amine compounds were investigated by gas chromatography (GC)-nitrogen phosphorous detector (NPD). To this end, three different amine compounds (methyl amine (MTA), dimethyl amine (DMA), and trimethyl amine (TMA)) were analyzed in a series of calibration experiments. As the first step, the liquid phase standards of these amines were calibrated at 5 concentration levels with the fixed standard volume (FSV) method. The results showed that response factors of three amines were in the order of TMA > DMA > MTA with their correlation coefficients (r) above 0.97. The results of FSV calibration also indicated that the analytical sensitivity of amines increased with reducing injection volume due to the decrease in matrix effect. However, when these compounds were calibrated by the fixed standard concentration (FSC) method, the results were not stable under most circumstances. In addition, the reliability of internal calibration approach was also tested using pyridine, nitrobenzene, and nicotine. When pyridine was used as internal standard of these amines, analytical reproducibility of amines was improved greatly compared to external calibration results (e.g., an improvement of 75% for DMA).

GC/NPD 방식에 기초한 아민류 악취물질의 분석특성 비교

김용현,안지원,김기현

초록