Journal Search Engine
Download PDF Export Citation Korean Bibliography PMC Previewer
ISSN : 1598-6616(Print)
ISSN : 2287-6731(Online)
Journal of Korean Society of Odor Research and Engineering Vol.9 No.1 pp.6-15
 

고체상 미량추출법(SPME)과 GC-MS에 기초한 신규악취물질들의 분석방법

Abstract

In this study, the calibration characteristics of six new odorant VOCs along with two reference compound were evaluated based on GC-MS (SIM). For the proper evaluation of the GC/MS method, calibration of 7 VOCs were made both by the SPME and direct injection (DI). The minimum detection limit (MDL) of those VOC determined by SPME method fell in the range of 0.31∼0.38 ppb while those of DI from 6.18∼14.3 ppb. The analytical performance of our SPME/GC-MS method was compared against those tested previously by other study groups. Although more efforts are needed to establish the reliability of our SPME/GC-MS method, the method adapted in this study can be recommended as one of the standard methods for the analysis of the target VOCs in this study to settle formal experiment methods.

고체상 미량추출법(SPME)과 GC-MS에 기초한 신규악취물질들의 분석방법

이기한,수딜 팬디,김기현

초록