Journal Search Engine
Download PDF Export Citation Korean Bibliography PMC Previewer
ISSN : 2288-9167(Print)
ISSN : 2288-923X(Online)
Journal of Odor and Indoor Environment Vol.13 No.1 pp.15-19
DOI : https://doi.org/10.15250/joie.2014.13.1.15

저농도 암모니아 측정을 위한 대한 고감도 검출기기 연구

안민호1, 백남현1, 황수현, 조영민*
1에이씨케이(주)
경희대학교 환경공학과

Highly sensitive sensing tube detector for low level ammonia

Young-Min Jo*, Min-Ho Ahn1, Nam-Hyun Baek1, Soo-Hyun Hwang
Department. of Environmental Science & Engineering, Kyunghee University, Kyunggido, Korea
1ACK Co., Ltd.
(Received 26 December, 2013; Revised 11 March, 2014; Accepted 25 March, 2014)

Abstract

Ammonia is one of typical index substances of environmental odor. This study describes a rapid and simpleadsorption tube type detector to be able to measure the level of strict emission standards with quantitativeconcentration of gaseous ammonia as the regulations are enhanced. a gas detector tube can be measured theammonia gas concentration of 0.23 ppm, 0.5 ppm, 1 ppm, 1.5 ppm and 2 ppm through the relation between theSampling rate Kog (=DA/L) and the discoloration reaction rate by physico-chemical analysis. Also this study canbe predicted the relation between the sampling volume and the sampling time. and it can be designed to measuringthe range of various concentrations. In this case study, an optimum design was found with an sampling rate, 42ml/min and with a tube diameter, 3.0 mm. Adsorption and diffusion rate was evaluated by Fick's law.

Reference